Pushing a compact 15 cm long ultra-high resolution drift tube ion mobility spectrometer with R = 250 to R = 425 using peak deconvolution

Show simple item record

dc.identifier.uri http://dx.doi.org/10.15488/4407
dc.identifier.uri https://www.repo.uni-hannover.de/handle/123456789/4447
dc.contributor.author Kirk, Ansgar T. ger
dc.contributor.author Zimmermann, Stefan ger
dc.date.accessioned 2019-01-24T14:15:59Z
dc.date.available 2019-01-24T14:15:59Z
dc.date.issued 2015
dc.identifier.citation Kirk, A.T:; Zimmermann, S.: Pushing a compact 15 cm long ultra-high resolution drift tube ion mobility spectrometer with R = 250 to R = 425 using peak deconvolution. In: International Journal for Ion Mobility Spectrometry 18 (2015), Nr. 1-2, S. 17-22. DOI: https://doi.org/10.1007/s12127-015-0166-z ger
dc.description.abstract Recently, we have published an analytical model describing the resolution and signal-to-noise-ratio of peaks present in an ion mobility spectrum acquired by a drift tube instrument. Based on this model, we designed a combination of a compact 10 cm long drift tube and a fast transimpedance amplifier achieving a resolution above 180. Here, we present results using an improved drift tube with slightly longer drift length of 15 cm and higher drift voltages, resulting in a resolution of R = 250 for the positive reactant ion peak and R = 230 for the negative reactant ion peak. By applying a deconvolution algorithm based on the Jansson method, a two- to three-fold separation improvement is possible in theory. However, as a larger resolution improvement than the initial resolution loss by diffusion may lead to erroneous results, we limited the resolution enhancement to 70 %, reaching a resolution of R = 425 for deconvolved spectra. The final publication is available at Springer via https://doi.org/10.1007/s12127-015-0166-z. ger
dc.language.iso eng ger
dc.publisher Berlin, Heidelberg : Springer
dc.relation.ispartofseries International Journal for Ion Mobility Spectrometry 18 (2015), Nr. 1-2 ger
dc.rights Es gilt deutsches Urheberrecht. Das Dokument darf zum eigenen Gebrauch kostenfrei genutzt, aber nicht im Internet bereitgestellt oder an Außenstehende weitergegeben werden. ger
dc.subject Ion mobility spectrometry eng
dc.subject Drift tube eng
dc.subject Ultra-high resolution eng
dc.subject Deconvolution eng
dc.subject Jansson method eng
dc.subject.ddc 621,3 | Elektrotechnik, Elektronik ger
dc.title Pushing a compact 15 cm long ultra-high resolution drift tube ion mobility spectrometer with R = 250 to R = 425 using peak deconvolution eng
dc.type Article ger
dc.type Text ger
dc.relation.essn 1865-4584
dc.relation.issn 1435-6163
dc.relation.doi 10.1007/s12127-015-0166-z
dc.bibliographicCitation.firstPage 17
dc.bibliographicCitation.lastPage 22
dc.description.version acceptedVersion ger
tib.accessRights frei zug�nglich ger


Files in this item

This item appears in the following Collection(s):

Show simple item record

 

Search the repository


Browse

My Account

Usage Statistics