Electron capture detector based on a non-radioactive electron source: Operating parameters vs. Analytical performance

Zur Kurzanzeige

dc.identifier.uri http://dx.doi.org/10.15488/3332
dc.identifier.uri http://www.repo.uni-hannover.de/handle/123456789/3362
dc.contributor.author Bunert, Erik
dc.contributor.author Kirk, Ansgar T.
dc.contributor.author Oermann, Jens
dc.contributor.author Zimmermann, Stefan
dc.date.accessioned 2018-05-18T12:03:53Z
dc.date.available 2018-05-18T12:03:53Z
dc.date.issued 2017
dc.identifier.citation Bunert, E.; Kirk, A.T.; Oermann, J.; Zimmermann, S.: Electron capture detector based on a non-radioactive electron source: Operating parameters vs. Analytical performance. In: Journal of Sensors and Sensor Systems 6 (2017), Nr. 2, S. 381-387. DOI: https://doi.org/10.5194/jsss-6-381-2017
dc.description.abstract Gas chromatographs with electron capture detectors are widely used for the analysis of electron affine substances such as pesticides or chlorofluorocarbons. With detection limits in the low pptv range, electron capture detectors are the most sensitive detectors available for such compounds. Based on their operating principle, they require free electrons at atmospheric pressure, which are usually generated by a β- decay. However, the use of radioactive materials leads to regulatory restrictions regarding purchase, operation, and disposal. Here, we present a novel electron capture detector based on a non-radioactive electron source that shows similar detection limits compared to radioactive detectors but that is not subject to these limitations and offers further advantages such as adjustable electron densities and energies. In this work we show first experimental results using 1,1,2-trichloroethane and sevoflurane, and investigate the effect of several operating parameters on the analytical performance of this new non-radioactive electron capture detector (ECD). © Author(s) 2017. eng
dc.language.iso eng
dc.publisher Göttingen : Copernicus GmbH
dc.relation.ispartofseries Journal of Sensors and Sensor Systems 6 (2017), Nr. 2
dc.rights CC BY 4.0 Unported
dc.rights.uri https://creativecommons.org/licenses/by/4.0/
dc.subject Electron capture detector eng
dc.subject Gas chromatograph eng
dc.subject sensor system eng
dc.subject.ddc 621,3 | Elektrotechnik, Elektronik ger
dc.title Electron capture detector based on a non-radioactive electron source: Operating parameters vs. Analytical performance
dc.type Article
dc.type Text
dc.relation.issn 2194-8771
dc.relation.doi https://doi.org/10.5194/jsss-6-381-2017
dc.bibliographicCitation.issue 2
dc.bibliographicCitation.volume 6
dc.bibliographicCitation.firstPage 381
dc.bibliographicCitation.lastPage 387
dc.description.version publishedVersion
tib.accessRights frei zug�nglich