IMS Instrumentation I : Isolated data acquisition for ion mobility spectrometers with grounded ion sources

Zur Kurzanzeige

dc.identifier.uri http://dx.doi.org/10.15488/10862
dc.identifier.uri https://www.repo.uni-hannover.de/handle/123456789/10944
dc.contributor.author Lippmann, Martin
dc.contributor.author Kirk, Ansgar T.
dc.contributor.author Hitzemann, Moritz
dc.contributor.author Zimmermann, Stefan
dc.date.accessioned 2021-05-03T09:10:21Z
dc.date.available 2021-05-03T09:10:21Z
dc.date.issued 2020
dc.identifier.citation Lippmann, M.; Kirk, A.T.; Hitzemann, M.; Zimmermann, S.: IMS Instrumentation I : Isolated data acquisition for ion mobility spectrometers with grounded ion sources. In: International journal for ion mobility spectrometry : official publication of the International Society for Ion Mobility Spectrometry 23 (2020), S. 69-74. DOI: https://doi.org/10.1007/s12127-020-00260-5
dc.description.abstract The drift voltage required for operating ion mobility spectrometers implies high voltage isolation of either the ion source or the detector. Typically, the detector is grounded due to the sensitivity of the small ion currents to interferences and thus higher requirements for signal integrity than the ion source. However, for certain ion sources, such as non-radioactive electron emitters or electrospray ionization sources, or for coupling with other instruments, such as gas or liquid chromatographs, a grounded ion source is beneficial. In this paper, we present an isolated data acquisition interface using a 16 bit, 250 kilosamples per second analog to digital converter and fiber optic transmitters and receivers. All spectra recorded via this new data acquisition interface and with a grounded ion source show the same peak shapes and noise when compared with a grounded detector, allowing additional freedom in design. © 2020, The Author(s). eng
dc.language.iso eng
dc.publisher Berlin ; Heidelberg : Springer
dc.relation.ispartofseries International journal for ion mobility spectrometry : official publication of the International Society for Ion Mobility Spectrometry 23 (2020)
dc.rights CC BY 4.0 Unported
dc.rights.uri https://creativecommons.org/licenses/by/4.0/
dc.subject ion mobility spectrometer eng
dc.subject data aquisitation eng
dc.subject amplifier eng
dc.subject isolated eng
dc.subject floating eng
dc.subject grounded ion source eng
dc.subject.ddc 530 | Physik ger
dc.title IMS Instrumentation I : Isolated data acquisition for ion mobility spectrometers with grounded ion sources
dc.type Article
dc.type Text
dc.relation.essn 1865-4584
dc.relation.issn 1435-6163
dc.relation.doi https://doi.org/10.1007/s12127-020-00260-5
dc.bibliographicCitation.volume 23
dc.bibliographicCitation.firstPage 69
dc.bibliographicCitation.lastPage 74
dc.description.version publishedVersion
tib.accessRights frei zug�nglich