High tilt susceptibility of the Scintrex CG-5 relative gravimeters

Zur Kurzanzeige

dc.identifier.uri http://dx.doi.org/10.15488/4960
dc.identifier.uri https://www.repo.uni-hannover.de/handle/123456789/5004
dc.contributor.author Reudink, R. ger
dc.contributor.author Klees, R. ger
dc.contributor.author Francis, O. ger
dc.contributor.author Kusche, J. ger
dc.contributor.author Schlesinger, R. ger
dc.contributor.author Shabanloui, Akbar ger
dc.contributor.author Sneeuw, N. ger
dc.contributor.author Timmen, L. ger
dc.date.accessioned 2019-06-17T12:04:36Z
dc.date.available 2019-06-17T12:04:36Z
dc.date.issued 2014
dc.identifier.citation Reudink, R. et al.: High tilt susceptibility of the Scintrex CG-5 relative gravimeters. In: Journal of Geodesy 88 (2014), Nr. 6, S. 617-622. DOI: https://doi.org/10.1007/s00190-014-0705-0 ger
dc.description.abstract We report on the susceptibility of the Scintrex CG-5 relative gravimeters to tilting, that is the tendency of the instrument of providing incorrect readings after being tilted (even by small angles) for a moderate period of time. Tilting of the instrument can occur when in transit between sites usually on the backseat of a car even using the specially designed transport case. Based on a series of experiments with different instruments, we demonstrate that the readings may be offset by tens of µGal. In addition, it may take hours before the first reliable readings can be taken, with the actual time depending on how long the instrument had been tilted. This sensitivity to tilt in combination with the long time required for the instrument to provide reliable readings has not yet been reported in the literature and is not addressed adequately in the Scintrex CG-5 user manual. In particular, the inadequate instrument state cannot easily be detected by checking the readings during the observation or by reviewing the final data before leaving a site, precautions suggested by Scintrex Ltd.In regional surveys with car transportation over periods of tens of minutes to hours,the gravity measurements can be degraded by some 10 µGal. To obtain high-quality results in line with the CG-5 specifications, the gravimeters must remain in upright position to within a few degrees during transits. This requirement may often be unrealistic during field observations,particularly when observing in hilly terrain or when walking with the instrument in a backpack. The final publication is available at https://doi.org/10.1007/s00190-014-0705-0. ger
dc.language.iso eng ger
dc.publisher Berlin Heidelberg : Springer
dc.relation.ispartofseries Journal of Geodesy 88 (2014), Nr. 6 ger
dc.rights Es gilt deutsches Urheberrecht. Das Dokument darf zum eigenen Gebrauch kostenfrei genutzt, aber nicht im Internet bereitgestellt oder an Außenstehende weitergegeben werden. ger
dc.subject Scintrex CG-5 autograv eng
dc.subject hysteresis eng
dc.subject tilt eng
dc.subject susceptibility to tilt eng
dc.subject accuracy degradation eng
dc.subject Scintrex CG5 Gravimeter ger
dc.subject Neigung ger
dc.subject Genauigkeit Degradierung ger
dc.subject.ddc 520 | Astronomie, Kartographie ger
dc.title High tilt susceptibility of the Scintrex CG-5 relative gravimeters eng
dc.type Article ger
dc.type Text ger
dc.relation.essn 1432-1394
dc.relation.isbn 10.1007/s00190-014-0705-0
dc.relation.issn 0949-7714
dc.bibliographicCitation.firstPage 617
dc.bibliographicCitation.lastPage 622
dc.description.version acceptedVersion ger
tib.accessRights frei zug�nglich ger


Die Publikation erscheint in Sammlung(en):

Zur Kurzanzeige

 

Suche im Repositorium


Durchblättern

Mein Nutzer/innenkonto

Nutzungsstatistiken