Traceability of the Hannover FG5X-220 to the SI units

Zur Kurzanzeige

dc.identifier.uri http://dx.doi.org/10.15488/4153
dc.identifier.uri https://www.repo.uni-hannover.de/handle/123456789/4187
dc.contributor.author Schilling, Manuel ger
dc.contributor.author Timmen, Ludger ger
dc.date.accessioned 2018-12-12T09:06:26Z
dc.date.available 2018-12-12T09:06:26Z
dc.date.issued 2015-06-22
dc.identifier.citation Schilling, M.; Timmen, L.: Traceability of the Hannover FG5X-220 to the SI units. Poster, 26th IUGG General Assembly, 22.06.-02.07.2015, Prag, Tschechische Republik. http://iugg.org/archive/iugg2015prague/www.iugg2015prague.com/index.html ger
dc.description.abstract The absolute measurement of g is currently realized through the laser interferometric measurement of a free falling retro-reflector. The Micro-g LaCoste FG5X is a free-fall gravimeter with a laser interferometer in Mach-Zehnder configuration which uses simultaneous time and distance measurements to calculate the absolute value of g. Because the instrument itself contains the necessary working standards for precise time and length measurements, it is considered independent of external references. The timing is kept with a 10 MHz rubidium oscillator with a stability of 5×10e−10 . The length unit is realized by the laser interferometer. The frequency calibrated and iodine stabilized helium-neon laser has a wavelength of 633 nm and an accuracy of 2.5×10e−11. In 2012 the FG5-220 of the Institut für Erdmessung (IfE) was upgraded to the FG5X-220. The upgrade included a new dropping chamber with a longer free fall and new electronics including a new rubidium oscillator. The metrological traceability to measurement units of the Système International d’unités (SI unit) is ensured by two complementary and successive approaches: the comparison of frequencies with standards of higher order and the comparison of the measured g to a reference measured by absolute gravimeters defined as primary standards within the SI. A number of experiments to test the rubidium oscillator were performed. The oscillator showed a linear drift of 0.2×10e−3 Hz per month (=0.3 nm/s² per month) in the first 18 months of use. A jump in the frequency of 0.01 Hz (=20 nm/s² ) was revealed recently and the drift rate changed to −0.5×10e−3 Hz /month. ger
dc.language.iso eng ger
dc.rights CC BY-NC 3.0 DE ger
dc.rights.uri http://creativecommons.org/licenses/by-nc/3.0/de/ ger
dc.subject absolute gravimetry eng
dc.subject frequency standard eng
dc.subject gravimeter comparison eng
dc.subject SI units eng
dc.subject Absolut Gravimetrie ger
dc.subject Frequenzstandard ger
dc.subject Gravimetervergleich ger
dc.subject SI Einheiten ger
dc.subject.ddc 550 | Geowissenschaften ger
dc.title Traceability of the Hannover FG5X-220 to the SI units eng
dc.type ConferenceObject ger
dc.type Text ger
dc.description.version publishedVersion ger
tib.accessRights frei zug�nglich ger


Die Publikation erscheint in Sammlung(en):

Zur Kurzanzeige

 

Suche im Repositorium


Durchblättern

Mein Nutzer/innenkonto

Nutzungsstatistiken