Minimally invasive mastoidectomy approach using a mouldable surgical targeting system. A proof of concept

Zur Kurzanzeige

dc.identifier.uri http://dx.doi.org/10.15488/10498
dc.identifier.uri https://www.repo.uni-hannover.de/handle/123456789/10575
dc.contributor.author Rau, Thomas S.
dc.contributor.author Witte, Sina
dc.contributor.author Uhlenbusch, Lea
dc.contributor.author Lexow, Jakob
dc.contributor.author Hügl, Silke
dc.contributor.author Kahrs, Lüder A.
dc.contributor.author Majdani, Omid
dc.contributor.author Lenarz, Thomas
dc.date.accessioned 2021-03-09T09:50:56Z
dc.date.available 2021-03-09T09:50:56Z
dc.date.issued 2018
dc.identifier.citation Rau, T.S.; Witte, S.; Uhlenbusch, L.; Lexow, J.; Hügl, S. et al. : Minimally invasive mastoidectomy approach using a mouldable surgical targeting system. A proof of concept. In: Current Directions in Biomedical Engineering 4 (2018), Nr. 1, S. 403-406. DOI: https://doi.org/10.1515/cdbme-2018-0096
dc.description.abstract Hearing restoration using a cochlear implant requires a surgical access to the inner ear. In order to enhance patient safety, reduce trauma, and shorten the patient's time under anaesthesia current research focusses on minimally invasive cochlear implantation surgery by drilling only a single bore hole. This demands a highly accurate surgical assistance device to guide the drill along a predetermined trajectory planned in patient's image data. In this study a recently developed surgical targeting system was evaluated for the first time in a human cadaver trial. After screwing a reference frame on a temporal bone specimen and imaging of both, a trajectory through the facial recess was planned in order to reach the middle ear. Based on this plan a patient specific surgical template including a linear guide for the surgical drill was composed utilizing bone cement. After the hardening of the bone cement the surgical template was mounted on top of the reference frame. The drilling could be performed as previously planned without harming facial nerve and chorda tympani. The deviation of the actual drill hole to the planned trajectory was 0.17 mm at the level of the facial recess. The minimal distance of the drill hole to the facial nerve was 0.59 mm. This proof-of-concept study demonstrates the feasibility of performing the access to the middle ear in a minimally invasive manner using the mouldable surgical targeting system. The presented process allows the patient specific individualization of a drill guide under sterile conditions. This might facilitate its integration into clinical routine. © 2018 Thomas S. Rau et al. eng
dc.language.iso eng
dc.publisher Berlin : Walter de Gruyter
dc.relation.ispartofseries Current Directions in Biomedical Engineering 4 (2018), Nr. 1
dc.rights CC BY-NC-ND 4.0 Unported
dc.rights.uri https://creativecommons.org/licenses/by-nc-nd/4.0/
dc.subject Cochlear implant eng
dc.subject Direct cochlear access eng
dc.subject Drill guide eng
dc.subject Drilling accuracy eng
dc.subject Image-guided surgery eng
dc.subject Micro-stereotactic frame eng
dc.subject Temporal bone eng
dc.subject.ddc 570 | Biowissenschaften, Biologie ger
dc.title Minimally invasive mastoidectomy approach using a mouldable surgical targeting system. A proof of concept
dc.type Article
dc.type Text
dc.relation.essn 2364-5504
dc.relation.doi https://doi.org/10.1515/cdbme-2018-0096
dc.bibliographicCitation.issue 1
dc.bibliographicCitation.volume 4
dc.bibliographicCitation.firstPage 403
dc.bibliographicCitation.lastPage 406
dc.description.version publishedVersion
tib.accessRights frei zug�nglich